Resistivity / Hall Effect Package
Abbess Instruments Resistivity / Hall Effect Probes Instrument package (IPACK-1) Voltage/Current Source Nanovoltmeter Relay Matrix |
|
leadless sample carrier |
|
| The Resistivity / Hall Effect probes designed and built to fit any ABBESS cryostat, to allow the customer to measure resistance changes as small as 10-6 ohm. The probe has a theoretical sensitivity of about 10-8 ohm:
The probes can accept samples up to the limits of the cryostat used. The probes typically has a minimum of 4 contacts, but may have up to 256. The Instrument Packages' (IPACK-1) contains a voltage/current source that supplies current from 100 micro amp to 10.1 amp with an accuracy of ±(0.l % + 0.1 mA) at the low end and ±(0.5% + 10 mA) at the high end. The IPACK-l's nanovoltmeter can read from 10nV to 1000V with accuracies of ±(0.006% + 5nV) at the low end and ±(0.005% + 1.5V) at the high end. The IPACK-1 can interface with the optional system controller(computer). |
| General Technical Specifications (Specifications may vary depending on final configuration) |
|
| Resistivity Measurement | Sensitivity 10-7ohms or better; theoretical sensitivity 10-9 ohms Uses quasi-DC measurement technique to avoid effects of sample inductance and capacitance while canceling thermal emf; complete measurements every 2 seconds over IEEE-488 bus Range 10-7 to 2 ohms (optional to 1000 ohms) Precision ±0.1%; accuracy (1 year) ±0.1% for R> 10-4 ohms; ±0.5% for 10-6< R < 10-4 ohms |
| Current Control | Minimum current is 0.1 nA Maximum current is 10.1 amp |
| Voltage Measurement | Recommended Nanovoltmeter: 10 nV resolution, 5.5-6.5 digit resolution, ranges to 1000 V, IEEE-488 interface |
| Current/Voltage source | Recommended Voltage/Current Source: 100 mA to 10 A or 1 V to 100 V full scale output with 3 digit resolution, 100 watts, IEEE-488 interface |
| Sample Environments (Dependent on Cryostat Used) | |
|---|---|
| Cryostat CTC-1 | Range: 4K to 925K Stability: Dependent on temperature controller selected Rate of Change: 300K to 77K in 5 minutes, 300K to 925K in 30 minutes Sample environment is isolated from cryogen chamber allowing user to choose gases or vacuum which will be in contact with the sample Sample Size: Up to .500 inch x 1.00 inch x .050 inch thick; max. electrical contacts 12 |
| Cryostat CTC-2 | Range: 1 0K to 1000K Stability: Dependent on temperature controller selected Rate of Change: 300K to 77K in 1 0 minutes, 300K to 800K in 30 minutes Sample environment is isolated from cryogen chamber allowing user to choose gases or vacuum which will be in contact with the sample Pressure: 10-6 Torr to 500 psig Sample Size: Up to 1.00 inch x 2.00 inch x .500 inch thick; max. electrical contacts 24 |
| Cryostat CTC-6 | Range: 4K to 425K Stability: Dependent on temperature controller selected Sample plate is copper cold finger to minimize gradients Cryostat is table or desk mountable Sample Size: Up to 1 inch x 1.25 inch; max. electrical contacts 256 |
Faraday and Hall Measurement System
Back to top of Abbess Main Home Page|
If you have any questions, need custom configurations, or a quotation,